Zero Calibration:
• Dip the probe into MA9070  zero oxygen
solution and stir gently for 2-3 minutes.
• Wait for the probe to reach it’s lowest
stable reading.
• Adjust the zero D.O. calibration
trimmer (located on the front
panel) until the display reads
"0.0".
Slope Calibration:
It is suggested to perform the slope calibration in water saturated
air.
• Rinse the probe with a large amount of clean water to
remove any residual zero oxygen solution.
• Dry the probe tip and allow a few minutes for the D.O. probe
to stabilize while suspending over a container of water in the
air.
• Press and hold the CAL key.
• Adjust the slope trimmer on the
front panel of the meter to read
"100%" on the LCD (while still
holding the CAL button).
• Release the CAL key and the
LCD will display a value in ppm
of oxygen.
The  zero calibration of the MW 600 is very stable, therefore
this procedure needs only to be performed whenever the probe
is replaced
. However, if most of the measurements are closer
to zero, more frequent zero calibration is advised.
Slope calibration can be easily performed on a weekly basis.