IBM 1000-90 ユーザーズマニュアル

ページ / 22
10
COMM.EXE
LI-1800 (UW) Portable Spectroradiometers
The LI-1800 and LI-1800UW Portable Spectroradiometers are
configured as DCE devices.  There are two ways to connect the
computer to the LI-1800 or LI-1800UW.
Terminal Port
1.
Connect the computer to the terminal port on the LI-1800 using
the 1000-04 printer cable, and use the COMM program as a
terminal emulator.
2.
Note that the LI-1800 does not recognize commands sent in
lowercase letters.
Output Port
Use the 2232S cable to connect the computer to the 9-pin output port
of the LI-1800.  You will need another terminal to control the
LI-1800; the COMM program will be used to capture data sent from
the output port.
LI-6200 Portable Photosynthesis System
The LI-6200 is configured as a DTE device.  To send files from the
LI-6200 to a computer, use the following functions:
FCT 51 - Set the I/O parameters as follows:
BAUD = 4800
DATA BITS = 8
STOP BITS = 2
PARITY = none
CHECK CD = N
CHECK DSR = N
ENQ/ACK HSHK = N (not used)
XON/XFF HSHK = N (not used)
XON CHAR = 11 (not used)
XOFF CHAR = 13 (not used)
FLAG PARITY = N (not used)
LOCAL ECHO = N (not used)
SOM =00 (not used)
EOL = 0D,OA
EOL WAIT = 0
EOM = 00 (not used)