NI 5422 Specifications
4
ni.com
Maximum Output Voltage
Maximum 
Output 
Voltage
Path
Load
Maximum Output Voltage (V)
The combination 
of Amplitude and 
Offset is limited 
by the Maximum 
Output Voltage.
Direct
50 
Ω
±0.500
1 k
Ω
±0.953
Open
±1.000
Low-
Gain 
Amplifier
50 
Ω
±1.000
1 k
Ω
±1.905
Open
±2.000
High-
Gain 
Amplifier
50 
Ω
±6.000
1 k
Ω
±11.43
Open
±12.00
Accuracy
DC Accuracy
For the Low-Gain or High-Gain Amplifier Path: 
±0.2% of Amplitude ± 0.05% of Offset ± 500 µV 
(within ±10 °C of self-calibration temperature)
±0.4% of Amplitude ± 0.05% of Offset ± 1 mV 
(0 °C to 55 °C)
All paths are 
calibrated for 
amplitude and 
gain errors. The 
Low-Gain and 
High-Gain 
Amplifier Paths 
also are 
calibrated for 
offset errors. 
Calibrated for 
high impedance 
load.
For the Direct Path:
Gain Accuracy: ±0.2% (within ±10 °C of self-calibration 
temperature)
Gain Accuracy: ±0.4% (0 °C to 55 °C)
DC Offset Error: ±30 mV (0 °C to 55 °C)
AC Amplitude 
Accuracy
±1.0% of desired Amplitude ± 1 mV
50 kHz sine 
wave.
Output Characteristics
Output 
Impedance
50 
Ω nominal or 75 Ω nominal, software-selectable.
Output 
Coupling
DC
Table 1.   (Continued)
Specification
Value
Comments