A&D GP-102K ユーザーズマニュアル

ページ / 98
 
 86
Display             Error code 
Description 
 
Memory type error 
Type of memory set in the function table and type of 
data stored are different. 
For details, refer to “11. DATA MEMORY”. 
EC, E00 
Communications error 
A protocol error occurred in communications.   
Confirm the format, baud rate and parity. 
  EC, E01 
Undefined command error 
An undefined command was received.   
Confirm the command. 
EC, E02 
Not ready 
A received command can not be processed.   
e.g.  The balance received a Q command, but not in the 
weighing mode. 
e.g. The balance received a Q command while  
processing a RE-ZERO command. 
Adjust the delay time to transmit a command. 
EC, E03 
Timeout error 
If the timeout parameter is set to “t-Up1”, the balance 
did not receive the next character of a command within 
the time limit of one second. 
Confirm the communication. 
EC, E04 
Excess characters error 
The balance received excessive characters in a command.
Confirm the command. 
EC, E06 
Format error 
A command includes incorrect data. 
e.g. The data is numerically incorrect. 
Confirm the command. 
EC, E07 
Parameter setting error 
The received data exceeds the range that the balance can accept. 
Confirm the parameter range of the command. 
Other errors 
If the errors described above can not be released or 
other errors are displayed, contact the local A&D dealer. 
Other symbol 
 
When this indicator blinks, automatic self calibration is 
required. The indicator blinks when the balance detects 
changes in ambient temperature. If the balance is not 
used for several minutes with this indicator blinking, the 
balance performs automatic self calibration. The blinking 
period depends on the operating environment. 
 
18-3  Asking For Repair 
If the balance needs service or repair, contact the local A&D dealer. 
The balance is a precision instrument. Use much care when handling the balance and observe 
the following when transporting the balance. 
  Use the original packing material. 
  Remove the weighing pan from the main unit.