A&D GP-100K ユーザーズマニュアル

ページ / 98
 
 40
9-4  Description of the Class “Environment, Display” 
Condition ( 
Cond 
Cond 0
 
This parameter is for sensitive response to the fluctuation of a weight value. 
Used for powder target weighing, weighing a very light sample or when 
quick response weighing is required. 
 
After setting, the balance displays FAST. 
 
Cond 2
 
This parameter is for stable weighing with slow response. Used to prevent a 
weight value from drifting due to vibration or drafts. 
After setting, the balance displays SLOW. 
Notes 
In automatic response adjustment, the response rate is selected 
automatically.  
With “Hold function (
Hold)” set to “ON (1)”, this item is used to set the 
averaging time. 
 
Stability band width ( 5t-b )   
This item controls the width to regard a weight value as a stable value. When the fluctuation per 
second is less than the parameter, the balance displays the stabilization indicator and outputs or 
stores the data. The parameter influences the “Auto print mode” 
5t-b 0
 
This parameter is for sensitive response of the stabilization indicator. Used 
for exact weighing. 
 
5t-b 2
 
This parameter ignores slight fluctuation of a weight value. Used to prevent 
a weight value from drifting due to vibration or drafts. 
Note 
With “Hold function (
Hold)” set to “ON (1)”, this item is used to set the 
stabilization range. 
 
Hold function ( Hold ) (Animal weighing mode) 
This function is used to weigh a moving object such as an animal.   
When the weighing data is over the weighing range from zero and the display fluctuation is within the 
stabilization range for a fixed period of averaging time, the processing indicator illuminates and the 
balance displays the average weight of the animal. When the animal is removed from the weighing 
pan, the display returns to zero automatically.   
This function is available only when the hold function parameter is set to “1” (the animal mode 
indicator ANIMAL illuminates) and any weighing unit other than the counting mode is selected. 
The stabilization range and averaging time are set in “Condition (Cond)” and “Stability band width (5t-b)”. 
Weighing range 
Averaging time 
Stabilization range
GP-12K/20K 
Cond 0
5t-b 0 
Cond 1
5t-b 1 
GP-30K/30KS/32K/32KS/ 
40K/61K/61KS 
20 g 
Cond 2
2 seconds     Faster 
4 seconds        | 
8 seconds    More accurate 
5t-b 2 
Small
Big 
GP-60K/60KS/100K/100KS 
 
 
 
 
GP-102K 
50 g