Canon TX-20 Full Auto Tonometer マニュアル

ページ / 100
42
4BasicOperation
8
Press the LIMITER button to set the position where the nozzle stops. 
 CAUTION 
Be sure to set the position where the nozzle stops for each patient 
Look from the side of the patient when setting the stop position to see that the patient’s forehead 
and chin are firmly against the forehead rest and chin rest. If the stop position is not set correctly, 
the nozzle may contact the patient’s eye and result in injury.
1)  Check that the LIMITER OFF mark is displayed. 
LIMITER 
OFF mark 
2)  Slowly move the operation lever so the nozzle approaches the eye being examined. 
 
Look at the patient from the side to make sure that the patient’s forehead and chin are in secure 
contact with the forehead rest and chin rest.
3)  Set the distance between the nozzle and the patient’s eye to approximately 8 to 10 mm, and 
press the LIMITER button. 
 
The LIMITER OFF mark disappears and the nozzle stop position mark is displayed. 
Nozzle stop 
position mark 
StopdoingmeasurementsiftheLIMITEROFFmarkdoesnotdisappeareveniftheLIMITER
buttonispressed
Stop using the instrument, and contact your sales representative or local Canon dealer. 