Canon galvano gc-251 ユーザーズマニュアル

ページ / 88
 
CANON    Digital Galvano Scanner System
 
KP-1SM Series + GC-251 
Users Manual 1.00 
 
 
 
  
80
Parameter ID 
64 
Parameter 
Name 
DSP Operation Setting 
Data 
DSP operation setting (Unit: none) 
Return Value 
0: Parameter setting successful 
1: Parameter setting unsuccessful 
Explanation 
This parameter is used to set the activation mode. Each bit has a meaning. 
 
Bit 0    0: No automatic encoder correction at homing to the origin 
            1: Automatic encoder correction at homing to the origin 
 
Bit 1    0: Activation in internal clock mode 
    1: Activation in high-speed serial communication (external clock) mode 
 
Bit 2    0: High-speed serial communication specification (XY2-100) (Usually 0) 
 
Note: The setting is decimal. 
Related 
Command 
/Parameter 
 
 
Parameter ID 
66 
Parameter 
Name 
High Speed Serial Status Format 
Data 
High Speed Serial Status Format (Unit: none) 
Return Value 
0: Parameter setting successful 
1: Parameter setting unsuccessful 
Explanation 
The content of status output to the status line of high-speed serial communications can 
be selected. 
        1:  Controller Condition 
        2:  Current position (Axis 1) 
        3:  Current position(Axis 2) 
        4:  Current position(Axis1 , Axis2) 
(For details, see 2.7.2, “High Speed Serial Communications”) 
 
Related 
Command 
/Parameter 
 
 
Parameter ID 
67 
Parameter 
Name 
High Speed Serial Data Length 
Data 
High Speed Serial Data Length (Unit: bit)      Range = 16∼20 
Return Value 
0: Parameter setting successful 
1: Parameter setting unsuccessful 
Explanation 
Position Data length of High speed serial communication can be specified. 
Range =  16bit ∼  20bit 
(For details, see 2.7.2, “High Speed Serial Communications”) 
Related 
Command 
/Parameter