Important safeguards/notices for use in the 
medical environments
1. All the equipment connected to this unit shall be 
certified according to Standard IEC60601-1, 
IEC60950-1, IEC60065 or other IEC/ISO Standards 
applicable to the equipment.
2. When this unit is used together with other equipment 
in the patient area*, the equipment shall be either 
powered by an isolation transformer or connected via 
an additional protective earth terminal to system 
ground unless it is certified according to Standard 
IEC60601-1 and IEC60601-1-1.
3. The leakage current could increase when connected 
to other equipment.
4. The operator should take precautions to avoid 
touching the rear panel input and output circuitry and 
the patient at the same time.
5. Model LMD-2140MD is a monitor intended for use 
in a medical environment to display pictures from 
cameras or other systems. These models are 
nonpatient care equipment with respect to the 
requirement of leakage current.
Caution
When you dispose of the unit or accessories, you must 
obey the law in the relative area or country and the 
regulation in the relative hospital.
Symbols on the unit
R1.5m
* Patient Area
Symbol
Location
This symbol indicates
Bottom
Main power switch.
Press to turn the monitor 
on.
Bottom
Main power switch.
Press to turn the monitor 
off.
Bottom
The equipotential 
terminal which brings the 
various parts of a system 
to the same potential.
Bottom
Functional earth terminal
8
Front
Brightness control
6
Front
Contrast control
-
Front
Key inhibit
The setting are locked so 
that they cannot be 
changed.
Warning on power connection
Use a proper power cord for your local power supply.
The United States
Canada
Continental Europe
Plug type
HOSPITAL GRADE
E41395
E41395-A
HOSPITAL GRADE
LL33182
LL76662
LP-34A
Female end
LS-60
Cord type
H05VV-F
Rated Voltage & Current
10A/125V
10A/125V
10A/250V
Safety approval
UL
CSA
VDE