Toshiba lf546 ユーザーズマニュアル

ページ / 157
 
 
 65 − 
 
6 F 8 A 0 9 3 4  
 
Switch operation 
Display example 
Description 
STEP4(=END) 
 
 
 
CS UNIT 
 
   XXX/ZZ
Z
 
 
 
SET OK?
 
 
 
Pushing      when the cursor is positioned 
on the 7th character shows a message to 
confirm whether the setting is OK or not. 
If OK, push       . If you want to redo the 
setting, push       . 
Pushing      cancels the setting operation 
and exits the setting screen. 
 
Note :    The selectable characters are displayed cyclically as shown below: 
 
Symbol 1 
!  ”  #  $  %  &  ’  (  )  *  +  ,  −  .  / 
 
 
 
 
↓ 
Numeric 
characters 
0∼9 
 
 
 
 
↓ 
Symbol 1 
:  ;  <  =  >  ?  ・ 
 
 
 
 
↓ 
Alphabetical 
characters 
(uppercase) 
A∼Z 
 
 
 
 
↓ 
Alphabetical 
characters 
(lowercase) 
a∼z 
 
 
 
 
↓ 
Special 
character 
“  ”(Space) 
 
 
8.2.9 Span (Range)   
 
You can set the following constants in this setting item: 
 
(1) Range type 
(2) Unit of span (can be changed in Range 1) 
(3) Span 
(4) Hysteresis 
 
● Range type 
Multiple ranges can be used by selecting the range type. You can select a single range, 
multiple ranges, or forward/reverse multiple ranges. 
 
Select one from five types shown below: 
 
Selection items (display)
Description 
   SINGLE 
Single range 
      4F-0R 
Unidirectional flow,   
automatic selection of multiple ranges 
      2F-2R 
Bidirectional flows,   
automatic selection of multiple ranges 
      EXT 2F-0R 
Unidirectional flow,   
multiple ranges selected by external signal
      EXT 2F-2R 
Bidirectional flows,   
multiple ranges selected by external signal
N O
O K
ESC 
O K
N O
ESC