Model 5100-XX-IT Toxic Gas Sensor Module
 
Page:  25 
 
5.3.1  SENSOR EXPOSURE TO GAS 
Calibration gas must be delivered to the sensor using the flow rate and duration listed in below: 
Model Gas 
Flow 
Period 
5100-03-IT  Oxygen 
300 cc/min 
Until Stable 
5100-04-IT  Carbon Monoxide 
300 cc/min 
Until Stable 
 
5100-05-IT  Hydrogen Sulfide 
300 cc/min 
Until Stable 
5100-06-IT  Chlorine 
300 cc/min 
  Until Stable 
Function
Display
Description
Reference
M
E
S T
M
E
S T
M
E
S T
M
E
S T
M
E
S T
M
E
S T
M E
S
T
M E
S
T
M
E
S T
M
E
S T
M
E
S T
M
E
S T
Sub-Routines
XXXPPM
Default Display
Operation:  Confirm area clear of gas, or apply  zero air to 
sensor.
10PPM-SP
Banner:  Select  span, enter <E> to calibrate sensor
Enter
ACK
Zero gas setting acknowledged
C 4 PPM
Banner:  Apply span gas, then enter <E> to calibrate gas sensor
C 4 PPM
Operation:  Apply calbration gas.
Enter
10PPM-SP
Banner:  Select  span, enter <E> when done
Up
25PPM-SP
Operation:  Change Span Gas Value to 25PPM
Sub A
WAIT-300
Operation:  Five minute time out before sensor is returned to 
service.
Up
Adj-SPAN
Operation:  User adjustable value
Enter
25PPM-SP
Operation:  Ready for user adjustment
Mode Function - Calibrate
Enter
CAL-0PPM--
Banner:  Apply zero gas, enter <E> when done
Mode
CALIB:--
Sub A
Sub B
Key
Mode
ALMRSET:
Mode Function -  Alarm Reset
Table 5-1
Calibration
Banner:  Calibration Aborted
Mode
XXXPPM
Default Display
CXXXPPM
Operation:  As gas is applied the reading will increase - wait 3 
minutes
Enter
CAL-OK
Calibration Passed - now remove gas
WAIT-300
Operation:  Five minute time out before sensor is returned to 
service.
Enter
CAL-FAIL
Operation:  No calibration gas applied, or sensor did not respond 
correctly.
(Any)
Operation:  Hold magnet over Mode Switch for ten senconds to 
abort calibration
Sub C
Enter
C 25PPM
Banner:  Apply 25PPM gas, then enter <E> to calibrate gas 
sensor
Sub B
C 25PPM
Operation:  Apply calbration gas.