Agilent Technologies ac power solutions 6843a ユーザーズマニュアル

ページ / 187
Language Dictionary - 3
43
MEASure:CURRent:AC?
FETCh:CURRent:AC?
Phase Selectable
These queries return the ac component rms current being sourced at the output terminals.
Query Syntax
MEASure:[SCALar]:CURRent:AC?
FETCh:[SCALar]:CURRent:AC?
Parameters
None
Examples
MEAS:CURR:AC?    FETC:CURR:AC?
Returned Parameters
<NR3>
Related Commands
MEAS:VOLT:AC?    MEAS:CURR?
MEASure:CURRent:ACDC?
FETCh:CURRent:ACDC?
Phase Selectable
These queries return the ac+dc rms current being sourced at the output terminals.
Query Syntax
MEASure:[SCALar]:CURRent:ACDC?
FETCh:[SCALar]:CURRent:ACDC?
Parameters
None
Examples
MEAS:CURR:ACDC?    FETC:CURR:ACDC?
Returned Parameters
<NR3>
Related Commands
MEAS:VOLT:ACDC?    MEAS:CURR:AMPL:MAX?
MEASure:CURRent:AMPLitude:MAXimum?
FETCh:CURRent:AMPLitude:MAXimum?
Phase Selectable
These queries return the absolute value of the peak current as sampled over one measurement
acquisition of 4096 data points.
Query Syntax
MEASure:[SCALar]:CURRent:AMPLitude:MAXimum?
FETCh:[SCALar]:CURRent:AMPLitude:MAXimum?
Parameters
None
Examples
MEAS:CURR:AMPL:MAX?    FETC:CURR:AMPL:MAX?
Returned Parameters
<NR3>
Related Commands
MEAS:CURR:ACDC?    MEAS:CURR:CRES?