PULSe:HOLD
This command specifies whether the pulse width or the duty cycle is to be held constant when the pulse
period changes. The following tables describe how the duty cycle, period, and width are affected when
one, two, or all three parameters are set in a single program message.
Command Syntax
 [SOURce:]PULSe:HOLD <parameter>
Parameters
WIDTh | DCYCle
*RST Value
WIDTh
Examples
PULS:HOLD DCYC
Query Syntax
[SOURce:]PULSe:HOLD?
Returned Parameters
<CRD>
Related Commands
PULS:COUN     PULS:DCYC     PULS:PER     PULS:WIDT
PULSe:HOLD = WIDTh
Parameter Set
Action
DCYCle
PERod
WIDTh
Set
Sets WIDTh. If WIDTh < PERiod, recalculates DCYCle; otherwise
recalculates the PERiod and DCYCle.
Set
Sets PERiod. If WIDTh < PERiod, recalculates DCYCle; otherwise
recalculates the PERiod and DCYCle.
Set
Set
Sets WIDTh. If WIDTh < PERiod, sets the PERiod and
recalculates DCYCle; otherwise recalculates the PERiod and
DCYCle.
Set
Sets DCYCle and recalculates the PERiod.
Set
Set
Sets DCYCle and WIDth and recalculates the PERiod.
Set
Set
Sets DCYCle and PERiod and recalculates the WIDTh.
Set
Set
Set
Sets WIDTh. If WIDTh < PERiod, sets the PERiod and
recalculates DCYCle; otherwise recalculates the PERiod and
DCYCle.
PULSe:HOLD = DCYCle
Parameter Set
Action
DCYCle
PERod
WIDTh
Set
Sets WIDTh and recalculates the PERiod.
Set
Sets PERiod and recalculates the WIDTh.
Set
Set
Sets WIDTh. If WIDTh < PERiod, sets the PERiod and
recalculates DCYCle; otherwise recalculates the PERiod and
DCYCle.
Set
Sets DCYCle and recalculates the PERiod.
Set
Set
Sets DCYCle and WIDth and recalculates the PERiod.
Set
Set
Sets DCYCle and PERiod and recalculates the WIDTh.
Set
Set
Set
Sets WIDTh. If WIDTh < PERiod, sets the PERiod and
recalculates DCYCle; otherwise recalculates the PERiod and
DCYCle.