Philips m5085a ユーザーズマニュアル
Ph
ilips
Med
ical Sys
tem
s
9
Running a Training Scenario
To run the selected training scenario, treat the Trainer exactly as you would
the HS1 Defibrillator.
1.
Pull the handle or press the green On/Off button to turn on the Trainer.
2.
Remove the clear
protective lid from the
training pads cartridge
and set it aside. Follow
the HeartStart Trainer’s
voice instructions.
3.
If using a clothed
manikin, remove the
clothing as instructed.
4.
Pull the tab at the top
of the training pads
cartridge to peel off
the resealable film
cover. Inside are two
adhesive pads on a
plastic liner. Lift the
pads out of the
cartridge and unwind the wires from the well. Peel one pad off the liner.
NOTE: Avoid touching the pad’s adhesive gel, because this may
interfere with the sensing circuitry of the Trainer. (Touching the gel does
not affect operation of the HeartStart Defibrillator.)
5.
Place the pad on the manikin.
Press the pad down firmly. Then
peel and place the other pad.
PRECAUTION: Do not place the
training pads on a person.*