GE PT878GC ユーザーズマニュアル

ページ / 308
Servicing the PT878GC
9-19
April 2004
Programming the Thickness Gauge (cont.)
6. The final value, the detection threshold, represents the percent 
of peak the PT878GC uses to make measurements. It will 
consider anything above the entered percentage as part of the 
signal. To change this value:
a. Press 
[ENTER]
 to open the text box.
b. Use the numeric keys to enter the recommended value.
c. Press 
[ENTER]
 to confirm your entry.
You have finished entering values for the Programming window. 
To leave this window:
Press the [
W
] arrow key to move to the Materials tab.
Press 
[F2]
 (Cancel) to return to Operate Mode without 
confirming the changes. The PT878GC returns to Operate 
Mode.
Press 
[F3]
 (OK) to confirm the new data and return to Operate 
Mode. The PT878GC returns to Operate Mode.