Hanna Instruments hi 9032 ユーザーズマニュアル

ページ / 33
19
18
Press the ATC/MTC key and the decimal
point will blink on the TEMP. display (both
when in the Celsius and in the Fahrenheit
scale).
Press the °C or
the  °F key and
the temperature
value will start
blinking.
Enter the temperature of the solution by
pressing the numeric keys and then press
ENTER key to confirm the value. The stored
value will be displayed and the decimal
point will stop blinking.
E.g. If you want to enter 37.2°C as tem-
perature value, proceed as follows:
Press 3
Press 7
Press .
Press 2
Press the ENTER key and the display will
show "37.2°C".
To clear the numeric entry,
press the CLEAR key.
The measurement reading displayed on
the large LCD will be manually compen-
sated for temperature variations.
ATC
MTC
0
°F
2
°C
3
°C
3
Cond.
7
%TC
.
°F
2
ENTER
PGM
Clear
To clear the numeric entry,
press the CLEAR key.
See page 45 for further information about
the TDS factor.
HI9032  is an auto-ranging TDS meter,
and the reading automatically switches from
one resolution to the next (from 0.00mg/l
up to 19.99 g/l).
When using HI 7686 together with HI 7669/
2W 
all readings are automatically compen-
sated for any temperature difference. Wait
for a few minutes for the temperature sen-
sor to attain thermal equilibrium with the
test solution before taking the measure-
ment. If the temperature of the conductivity
probe  and the solution is drastically apart,
a longer time should be allowed before tak-
ing readings.
Once the measurement reading stabilizes
the measurement is complete.
If further measurements are desired, rinse
the probe with a small amount of the next
solution and test the sample.
With Manual Temperature Compensation
Immerse the HI7686 probe in the solution
submerging the holes of the sleeve into
the liquid. Tap the probe lightly on the
bottom of the recipient to remove any air
bubbles which may have being trapped
inside the sleeve.
Take the temperature of
the solution with a
Checktemp or a glass
thermometer.
PGM
Clear