Hanna Instruments hi 9032 ユーザーズマニュアル

ページ / 33
13
12
Conductivity measurement mode
By pressing the Cond. key, the display
will show "CONDUCTIVITY" to confirm the
measurement mode.
The meter will automatically default the
temperature coefficient % value to 2.0%
(shown on the secondary display) every
time it is turned on.
Change the temperature coefficient to the
value of the solution, by pressing the %TC
key. The % value will start blinking on the
secondary LCD.
Enter the desired value by pressing the
numeric keys and then press the ENTER
key to confirm the value.
The stored value will be displayed and it
will stop blinking.
E.g. If you want to enter 2.5% as tem-
perature coefficient value, proceed as fol-
lows:
Press the %TC key
Press 2
Press .
Press 5
Cond.
7
CONDUCTIVITY
%TC
S
m
%TC
.
CONDUCTIVITY
%TC
S
m
Switch the instrument on
by pressing the ON key.
The meter automati-
cally defaults to the
conductivity mea-
surement mode with
2.0% as temperature
coefficient (on the
large LCD) and the
temperature in the
Celsius scale (on the TEMP. LCD).
Select a different measurement mode by
pressing the following keys:
the Fert. key for fertilizer
#1
the Fert. key twice for fer-
tilizer #2
the NaCl key for NaCl of
sea water saturation.
the T.D.S. key for total
dissolved solids
the Cond. key to enter the
conductivity measurement
mode again.
ON
CONDUCTIVITY
%TC
S
m
NaCl
9
Cond.
7
T.D.S.
4
Fert.
8
Fert.
8