Hanna Instruments hi 983321-0 ユーザーズマニュアル

ページ / 3
Instruction Manual
HI 983321-0
HI 983321-1
Panel-Mounted
Indicators &
Controllers
These Instruments are in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction and appear-
WARRANTY
These meters are warranted for two years against defects in work-
manship and materials when used for their intended purpose and
The probe is warranted for a
. This warranty is limited to repair or replacement
Damages due to accident, misuse, tampering or lack of prescribed main-
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service depart-
ment and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank  you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the   directives.
PRELIMINARY EXAMINATION
HI 983321-0  and HI 983321-1 are conductivity  indicators
and controllers with a relay output designed for simplicity of
use in a wide range of  applications.
The models are panel mounted with membrane keypads on
the front panel and an easy-to-read LCD display.
All the meters automatically compensate for the temperature
variation. The probe is easy to clean and requires little
maintenance.
Measurements are highly accurate and the meters can be
calibrated at one point.
Power supply, wiring and selection are made via the plug-
in terminal blocks on the rear panel.
LED indicators on the front panel identify whether the
controller is in set or measurement mode, and if alarm is
active.
GENERAL DESCRIPTION
SPECIFICATIONS
HI 983321
Range
0.00 to 19.99 ppm
Resolution
 0.01 ppm
Accuracy (@ 20°C)
±2% f.s.
Setpoint
Adjustable through multiturn trimmer
Alarm Condition LED ON and alarm contact closed when
measured TDS value is higher than setpoint
Alarm Output
2-contact relay, no fuse protected.
5A, 240 VAC, 30 VDC
Probe
HI 7634-00 EC/TDS probe
TDS conversion factor
0.5
Temp.Compensation      Automatic from 5 to 50°C
        (41 to 122°F); ß=2%/ºC
Calibration
Factory calibrated
and manual, at one point through trimmer
Calibration Solution
recommended >13 ppm
(should be prepared by using HI 9034 TDS meter)
Power supply:
External
HI 983321-0
 12 VDC
HI 983321-1
  115/230 VAC; 50/60Hz
Dimensions
79 x 49 x 95 mm (3.1 x 1.9 x 3.7”)
Weight:
HI 983321-0
 140 g (4.96 oz)
HI 983321-1
 250 g (8.85 oz)
Recommendations for Users
Before using these products, make sure that they are entirely suitable for
the environment in which they are used. Operation of these instruments in
residential areas could cause unacceptable interferences to radio and TV
equipment. The metal band at the end of the probe is sensitive to electro-
static discharges. Avoid touching this metal band at all times. During operation,
ESD wrist straps should be worn to avoid possible damage to the probe by
electrostatic discharges. Any variation introduced by the user to the sup-
plied equipment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltage at the
measurement surface exceeds 24 VAC or 60 VDC. Use plastic beakers to
minimize any EMC interferences.
To avoid damage or burns, do not perform any measurement in microwave
ovens.
h t t p : / / w w w . h a n n a i n s t . c o m
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meter is supplied with:
• Instruction manual;
• Mounting brackets.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
CE DECLARATION
OF CONFORMITY