Allow 1 or 2 minutes for
the electrode to stabi-
lize.
pH is displayed on the
primary display and tem-
perature on the second-
ary one.
If readings are out of
range  "---" will appear
on the LCD display.
For pH 301 only:
Press the .00/.000 key to change the resolu-
tion from two to three decimal points.
It is recommended that the electrode is rinsed
thoroughly for better conditioning. For this
rinsing process, it is recommended to use a
liberal amount of the sample to be measured.
If measurements are taken in different samples
successively, it is recommended that the elec-
trode and the temperature probe are rinsed
thoroughly with deionized water or distilled
water (if not available, tap water may also be
used) and then blot dry between each mea-
surement.
The pH reading is directly effected by tem-
perature. In order for the meter to measure
the pH accurately, temperature must be taken
into consideration.
A perfect equilibrium between the pH elec-
trode and the sample is reached in approxi-
mately 15 minutes.
To use the meter's Automatic Temperature
Compensation
 feature, submerge the
HI7669/2W temperature probe into the sample
°C
pH
pH
0.00/0.000
°C
pH
SETTING THE DATE AND THE TIME AND
THE LOGGING INTERVAL (for pH 302 only)
See program No. 1 (page 60).
VIEWING THE DATE AND THE TIME (for
pH 302 only)
Press TIME to display the time (HH/MM).
Press TIME again to display the date (MM/
DD).
TAKING pH  MEASUREMENTS
Make sure that the instrument has been cali-
brated for pH before taking pH measure-
ments (see page 29).
For pH 300 and pH 302 only:
Select the pH measurement mode
by pressing RANGE.
For pH 301 only:
Select the pH measurement
mode by pressing pH.
To take a pH measurement
simply submerge the tip (4cm/
1½") of the pH electrode and
the temperature probe into the
sample to be tested.
TIME
TIME
DATE
RANGE
pH
4 cm
(1½")