LifeSource ua-702 ユーザーズマニュアル

ページ / 20
Display
Symbol 
Recommended
Action
Remain still and refrain
from talking; measurement
in progress.
Press exhaust button to
release air.
Increase pressurization by
30 mmHg to 40 mmHg and
repeat measurement.
Replace all batteries 
with new ones.
Exhaust air from cuff, reap-
ply cuff properly, power up
again and repeat measure-
ment in quiet environment.
Exhaust air from cuff, 
reapply cuff properly, power
up again and repeat 
measurement.
Adjust exhaust rate to 2
mmHg to 5 mmHg.
Check for correct air con-
nector. Make sure air con-
nector is inserted correctly
into the hoses and monitor.
Condition/
Cause
Appears during measure-
ment; blinks and beeps
when pulse is detected.
Appears on display prior to
and during measurement.
Blinks if there is air in cuff
at power up.
Measurement begun, but
pressure is not sufficient for
reading.
Battery voltage is 
too low for monitor 
to work properly. 
Appears when blood pres-
sure cannot be measured
accurately due to erratic
pulse or noise interference.
Appears when pulse cannot
be measured accurately.
Occurs when exhaust 
velocity rate is more than 8
mmHg.
No Display Symbol
MEASURING
LOW BATTERY
ERROR
WHAT DISPLAY SYMBOLS MEAN
?
3
?
LifeSource Health Line (Toll-Free): 1-888-726-9966
ERROR
RE-MEASURE
PULSE ERROR
EXHAUST
SPEED FLASHING
CUFF HOSE
MONITOR
BULB  HOSE