6. TROUBLESHOOTING 
 
The LED indicator is off when switch is ON. 
Turn the switch on again. If LAM still does not go ON 
check that battery is installed properly and has sufficient 
charge. Replace battery if necessary.  
Spot shape or size has changed 
Emitter objective lens is dirty. Flush emitter lens with 
fresh water and wipe it with soft cloth. If this doesn’t 
help, send LAM for repair. 
No radiation when activating LAM with remote switch  
Turn the mode switch (13, Fig. 2) to OFF position and 
then turn LAM ON again. If this doesn’t help, replace the 
remote switch. 
Radiation doesn’t stop when remote switch is released  
Replace the remote switch. 
Moisture in the battery compartment. 
Check the O-ring on the battery compartment cover and 
replace it if necessary. If the battery compartment cover is 
damaged, send LAM for repair 
Shift of zero line when firing 
Send LAM for repair.