Chapter 3
VXI-MXI-2 Configuration and Installation
3-10
ni.com
When switch S3 is set so that the VXI-MXI-2 receives the SMB CLK10 
signal, you have the option to add a 50 
Ω termination to the signal by setting 
switch S4. S4 is unused—its setting does not matter—when S3 is 
configured to drive the external CLK SMB signal.
You can use an additional switch, S5, to control the polarity of the external 
CLK SMB signal when S3 is configured to drive it. S5 is unused—its 
setting does not matter—when S3 is configured to receive the external CLK 
SMB signal.
Figure 3-6 shows four configuration settings for the VXI-MXI-2. 
Figure 3-6a shows the default configuration, which is for driving the 
inverted external CLK SMB. Use the settings of Figure 3-6b to drive the 
noninverted external CLK SMB signal. Figure 3-6c illustrates the setting 
for receiving the external CLK SMB signal. Finally, you can configure the 
switches as shown in Figure 3-6d to receive the external CLK SMB signal 
with a 50 
Ω termination. 
Note
The settings of any switches shown with this pattern (
) have no bearing on the 
configuration described in any of the following figures.