National Instruments PCI-1200 ユーザーズマニュアル

ページ / 100
Chapter 5
Calibration
© National Instruments Corporation
5-3
Using the Calibration Function
The 
Calibrate_1200
 function and the 1200 Calibrate VI can either load 
the calibration DACs with the factory constants or the user-defined 
constants stored in the EEPROM, or you can perform your own calibration 
and directly load these constants into the calibration DACs. To use the 
Calibrate_1200
 function or the 1200 Calibrate VI for analog input 
calibration, ground an analog input channel at the I/O connector for offset 
calibration and apply an accurate voltage reference to another input channel 
for gain calibration. You should first configure the ADC for RSE mode, 
then for the correct polarity at which you want to perform data acquisition. 
To use the 
Calibrate_1200
 function or the 1200 Calibrate VI for analog 
output calibration, the DAC0 and DAC1 outputs must be wrapped back and 
applied to two other analog input channels. You should first configure the 
analog input circuitry for RSE and for bipolar polarity, then configure the 
analog output circuitry for the polarity at which you want to perform output 
waveform generation. 
Refer to your software documentation for more details on the 
Calibrate_1200
 function and the 1200 Calibrate VI.