Hanna Instruments hi 981411-0 ユーザーズマニュアル

ページ / 2
Instruction Manual
HI 981411-0
HI 981411-1
Panel-Mounted
pH Indicators &
Controllers
Manufactors since  1978
These Instruments are in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the de-
sign, construction and appearance of its products without
advance notice.
WARRANTY
HI 981411-0 and HI 981411-1 are warranted for two years against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the  .
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meter is supplied with:
 - Mounting brackets;
 - Instruction manual.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATION
HI 981411-0 and HI 981411-1 are pH indicators and
controllers with a relay output designed for simplicity of use
in a wide range of applications.
The models are panel-mounted with membrane keypads
on the front panel and an easy-to-read LCD display.
Power supply, wiring and selection are made via the plug-
in terminal blocks on the rear panel. The instruments are
equipped with a BNC socket and accept input from conven-
tional combination pH electrode.
LED indicators on the front panel identify whether the
controller is in set or reading mode, and if dosing device
is activated.
GENERAL DESCRIPTION
SPECIFICATIONS
   ISTR981411R5       06/2001
CE DECLARATION
OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the
environment in which it is used. Operation of this instrument in residential
areas could cause unacceptable interferences to radio and TV equipment.
The glass bulb at the end of the electrode is sensitive to electrostatic dis-
charges. Avoid touching this glass bulb at all times. During operation, ESD
wrist straps should be worn to avoid possible damage to the electrode by
electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade
the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24 VAC or 60 VDC. To avoid damages or burns,
do not perform any measurement in microwave ovens.
                    HI981411-0 and HI981411-1
14-4-99
HI 981411
Range
0.0 to 14.0 pH
Resolution
0.1 pH
Accuracy (@ 20°C/68°F)
±0.2 pH
Offset Calibration
Manual, through trimmer
Setpoint
Adjustable, from 0.0 to 14.0 pH
Dosing Selection
Acid or Alkaline, selectable on the back panel
Contact Open=Acid dosage=Relay ON if Measure > Setpoint
Contact Close=Alkaline dosage=Relay ON if  Measure < Setpoint
Power supply:
External
HI 981411-0
12 VDC
HI 981411-1
115/230 VAC; 50/60Hz
Dimensions
79 x 49 x 95 mm (3.1 x 1.9 x 3.7”)
Weight:
HI 981411-0
140 g (4.96 oz)
HI 981411-1
250 g (8.85 oz)
h t t p : / / w w w . h a n n a i n s t . c o m