eutech-instruments cadmium epoxy ユーザーズマニュアル

ページ / 20
Instruction Manual 
 
Cadmium Electrode
 
 
19
 
 
                
incorrect reference  
 
 
use recommended 
                        
 
filling solution     
 
 
filling solution 
 
Low Slope or             
standards contaminated     
 
prepare fresh  
No Slope             
 
or incorrectly made    
 
standards       
                      
                        
 
ISA not used     
 
 
use recommended ISA  
                                                
 
                    
 
 
standard used as ISA    
 
use ISA 
 
 
 
                  
electrode exposed       
 
check section 
                         
 
to complexing agents   
 
entitled 
Precipitation    
                              
 
 
 
                                  and Complexation
  
 
 
 
                
air bubble on   
             
remove bubble by 
                         
 
membrane                  
 
redipping probe 
 
 
"Incorrect Answer"     
incorrect scaling        
 
plot millivolts on  
(but calibration    
 
of semi-log paper        
 
the linear axis. On  
curve is good)                                
 
 
 
 
the log axis, be sure  
         concentration 
numbers 
         within 
each 
decade 
are 
 
                              
 
 
 
 
 
increasing with 
         increasing 
concentration. 
 
 
 
       
 
incorrect sign   
 
 
be sure to note sign of 
 
         millivolt 
number 
correctly 
 
 
                    
  
 
incorrect standards  
 
 
prepare fresh  standards 
 
 
 
               
wrong units used     
 
 
apply correct 
                                                
 
 
 
 
conversion factor: 
                                         
 
 
 
 
 
10
-3
M = 112 ppm as Cd
+2
 
 
 
 
 
                                              
     
 
             
complexing agents       
 
check section entitled 
 
  
       
 
in sample                                 
Precipitation and 
  
 
 
 
 
 
 
 
 
Complexation
use titration