eutech-instruments ecoscan do 6 do meter ユーザーズマニュアル

ページ / 76
Instruction Manual 
DO 6 
 
 
59
 
Setting 
Menu 
FUNCTION PARAMETER 
OPTING KEYS 
PARAMETER 
OPTIONS 
DEFAULT 
SETTING 
DPr.7 
Barometric Pressure 
Selection  Menu 
Parameter Title 
Only 
Parameter 
Title Only 
 
A.HG 
Selection of Barometric 
Pressure Setting 
Adjustment in mmHg 
▲ and ▼ 
500 to 1499 
mmHg 
760 mm 
Hg 
A.PA 
Selection of Barometric 
Pressure Setting 
Adjustment in Kilo Pascal 
▲ and ▼ 
66.6 to 199.9 
kPA 
101.3 kPA 
b.SAL 
Salinity Setting Adjustment 
▲ and ▼ 
0.0 to 50.0 ppt 
0.0 ppt 
 
Setting the Manual 
Temperature Compensation 
CAL, ▲ and ▼ 
0.0 to 50.0 °C 
25.0°C 
Note: The HOLD/ENTER key is used to confirm every parameter opted.