eutech-instruments lead epoxy ユーザーズマニュアル

ページ / 23
  
 
30 
  30.07 
  
 
40 
  31.07 
  
 
50 
  32.06 
 
If changes in temperature occur, the electrodes should be re-calibrated. The temperature range for 
the Eutech Lead Ion Electrode is 0
o
-80oC, provided that temperature equilibrium has occurred. If 
the temperature varies substantially from room temperature, equilibrium times up to one hour are 
recommended. 
 
Electrode Response 
 
Plotting the electrode mV potential against the lead concentration on semi-logarithmic paper results 
in a straight line with a slope of about 25 mV per decade.  (Refer to Figure 1.) 
The time needed to reach 99% of the stable electrode potential reading, the electrode response time, 
varies from several seconds in highly concentrated solutions to several minutes near the detection 
limit. 
 
A drifting potential reading or a decrease in electrode slope may mean that the electrode membrane 
needs polishing. 
 
To polish the membrane: 
 
1. 
If using polishing paper, cut off a 1-2" piece and place it face up on the lab bench. 
 
2. 
Put a few drops of distilled or deionized water in the center of the paper. 
 
3. 
Holding the paper (cotton) steady with one hand, bring the membrane of the electrode 
down perpendicular to the paper and, with a slight swirling motion, gently polish the tip of 
the electrode against the surface of the polishing paper (cotton) for a few seconds. 
 
4. 
Rinse the electrode surface with distilled or deionized water and soak the electrode tip in 
standard solution for about five minutes before use. 
 
5. 
If using jeweller's rouge, place a cotton ball on the table top and flatten it using the bottom 
of a beaker. 
 
6. 
Put 1-2 drops of distilled or deionized water in the center of the cotton pad. 
 
7. 
Add a small amount of jeweller's rouge to the damp cotton. 
 
8. 
Continue with Steps 3 and 4 above. 
 
Limits of Detection 
 
The upper limit of detection in pure lead perchlorate solutions is 0.1M. In the presence of other 
ions, the upper limit of detection is above 1.0x10
-2
M lead, but two factors influence this upper limit. 
Both the possibility of a liquid junction potential developing at the reference electrode and the salt 
extraction effect influence this upper limit. Some salts may extract into the electrode membrane at 
high salt concentrations, causing deviation from the theoretical response. Either dilute samples 
between 0.1M and 1.0x10
-2
M or calibrate the electrode at 4 or 5 intermediate points.