ZOOM BRUSH POSITION CHECK/ADJUSTMENT AND PATTERN CHECK
Note:
If dialog box of error code appears during the checking or adjustment, check the reason of error referring to page 4-26.
4-7-1. Zoom Brush Position Check
Equipment
• Personal Computer (PC)
• Camera DSLR-A100
• USB Cord With Connector
• Lens Adjustment Program
Note:
Lens Adjustment Program is downloadable from the ESI homepage.
1. Preparations
1)
Connect the checking lens to the camera.
2)
Start the lens adjustment program “LensAdjustment.exe” referring to “4-1-2. Lens Adjustment Program”.
Fig. 4-7-1
2. Checking Method
1)
Click the 
[Connect]
 button on the lens adjustment program.
Note:
Click the 
[End]
 button to disconnect the USB connection, then lens adjustment program will terminate.
Fig. 4-7-2
2)
Click the 
[Zoom]
 button on the lens adjustment program.
3)
Set the zoom position to Wide end, and check that the OK (Green) indicator of “Position” lights as shown in Fig. 4-7-3.
Note:
Lens focus position setting is not required.
Fig. 4-7-3
If the NG (Red) indicator of “Position” lights, perform the “4-7-2. Zoom Brush Position Adjustment and Pattern Check”.
Fig. 4-7-4
4)
Click the 
[Exit]
 button.
5)
Click the 
[End]
 button to terminate the lens adjustment program.
6)
Turn the POWER switch of the camera to OFF.