Insulet Corporation 017 ユーザーズマニュアル
126
10
Pod expiration advisory.
Change Pod now.
Advisory
Alarm
Alarm
Pod
and
PDM
and
PDM
2 sets of beeps repeat periodically.
Repeats every 15 minutes during
the last hour of the Pod’s life and is
independent of the Expiration noti-
fication that you can set. (See Chap-
ter 6, Using the Personal Diabetes
Manager.)
the last hour of the Pod’s life and is
independent of the Expiration noti-
fication that you can set. (See Chap-
ter 6, Using the Personal Diabetes
Manager.)
Escalates to Pod expired hazard
alarm.
alarm.
Press OK to acknowledge.
Change Pod. (See Chapter 5, Using the Pod.)
Pod expired.
Insulin delivery stopped.
Change Pod now.
Hazard
Alarm
Alarm
Pod
and
PDM
and
PDM
Continuous tone.
This alarm occurs when the Pod has
reached the end of its operating life.
Press OK to acknowledge.
reached the end of its operating life.
Press OK to acknowledge.
Change Pod.
Check blood glucose.
Stuck key detected.
Check Home button to
resolve.
resolve.
For technical support refer
to your User Guide.
to your User Guide.
Advisory
Alarm
Alarm
PDM
2 sets of beeps every 5 minutes.
Check the key described on the screen.
(In this example, the stuck key is the
Home button.) Press the key to unstick
it.
(In this example, the stuck key is the
Home button.) Press the key to unstick
it.
If this does not work, call Customer
Care.
Care.
Alarm
Type
Source
Beep Pattern
Action