Cleaning the MiniLink
Caution:  The charger and tester units are NOT water-tight. Do 
NOT immerse either in water.
For single-patient use
Wash your hands thoroughly. 
Dampen a clean cloth with mild liquid soap and warm water. Wipe the outside 
of the MiniLink.
Rinse the MiniLink under warm tap water but do NOT get water inside the 
connector.
Using an antibacterial hand-sanitizer (available at a local drugstore) on a clean, 
dry cloth, wipe down the MiniLink. Make sure to dry the inside of the connector.
Place the MiniLink on a clean, dry cloth and air dry for 2-3 minutes.
Caution:  DO NOT discard the MiniLink transmitter in a medical 
waste container or otherwise subject it to incineration. 
MiniLink contains a battery which may explode upon 
incineration.
For multiple-patient use
Wash your hands thoroughly.
Dampen a clean cloth with a mild liquid soap solution. Wipe the outside of the 
MiniLink. 
Rinse the MiniLink under warm tap water but do NOT get water inside the 
connector.
Apply 3-4 drops of a quaternary ammonium compound disinfectant (for 
example, Cavicide
®
) on a clean, dry cloth and wipe the MiniLink.
Hold the MiniLink and rinse with 70% Isopropyl alcohol. 
Place the MiniLink on a clean, dry, non-shedding cloth and air dry. 
Once the MiniLink is completely dry, place it in a sealed bag labeled with the 
cleaning date (see label example below).