Cleaning the MiniLink
Caution: The charger and tester units are NOT water-tight. Do
NOT immerse either in water.
For single-patient use
1 Wash your hands thoroughly.
2 Dampen a clean cloth with mild liquid soap and warm water. Wipe the outside
1 Wash your hands thoroughly.
2 Dampen a clean cloth with mild liquid soap and warm water. Wipe the outside
of the MiniLink.
3 Rinse the MiniLink under warm tap water but do NOT get water inside the
connector.
4 Using an antibacterial hand-sanitizer (available at a local drugstore) on a clean,
dry cloth, wipe down the MiniLink. Make sure to dry the inside of the connector.
5 Place the MiniLink on a clean, dry cloth and air dry for 2-3 minutes.
Caution: DO NOT discard the MiniLink transmitter in a medical
waste container or otherwise subject it to incineration.
MiniLink contains a battery which may explode upon
incineration.
MiniLink contains a battery which may explode upon
incineration.
For multiple-patient use
1 Wash your hands thoroughly.
2 Dampen a clean cloth with a mild liquid soap solution. Wipe the outside of the
1 Wash your hands thoroughly.
2 Dampen a clean cloth with a mild liquid soap solution. Wipe the outside of the
MiniLink.
3 Rinse the MiniLink under warm tap water but do NOT get water inside the
connector.
4 Apply 3-4 drops of a quaternary ammonium compound disinfectant (for
example, Cavicide
®
) on a clean, dry cloth and wipe the MiniLink.
5 Hold the MiniLink and rinse with 70% Isopropyl alcohol.
6 Place the MiniLink on a clean, dry, non-shedding cloth and air dry.
7 Once the MiniLink is completely dry, place it in a sealed bag labeled with the
6 Place the MiniLink on a clean, dry, non-shedding cloth and air dry.
7 Once the MiniLink is completely dry, place it in a sealed bag labeled with the
cleaning date (see label example below).