Vieworks Co. Ltd. FXRD-1012NAW ユーザーズマニュアル
Rev.1.0
Page 121 of 151
VW40-152-009
VIVIX-S 1012N User Manual
Self Diagnosis
You can perform self diagnosis of the detector or SCU from the Diagnosis window in the VIVIX Setup
program.
program.
1 Choose a desired item to diagnose.
2 Click Execute Test button located at the bottom of the Test Case window.
3 Check the progress and result of diagnosis for each item in the Result window.
4 Click Save to File to save the diagnosis result as a file when the self diagnosis is completed.
2 Click Execute Test button located at the bottom of the Test Case window.
3 Check the progress and result of diagnosis for each item in the Result window.
4 Click Save to File to save the diagnosis result as a file when the self diagnosis is completed.
Self diagnosis and resolution can be conducted by a user or a service engineer.
Sensitivity and calibration should be conducted by an authorized service engineer granted
by Vieworks.
by Vieworks.
Self-Diagnosis Items of Detector
Category
Item
Form
Expected problem
Action
Voltage
DDC
Decision
Poor condition of a tether
interface cable.
interface cable.
Poor power supply to the wired
operation mode.
operation mode.
Change a tether cable.
Contact a service engineer.
MAIN
Decision
Poor power supply to the
processor.
processor.
Contact a service engineer.
SIGNAL
Decision
Poor power supply to FPGA
Contact a service engineer.
Battery
Detection
Decision
Unattached battery
Defective ciruit connected to
a battery pack.
Check if a battery is inserted.
Contact a service engineer.