Vieworks Co. Ltd. FXRD-1012NAW ユーザーズマニュアル

ページ / 151
 
 
Rev.1.0
 
Page 130 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
8.1 
Trobleshooting 
When you encounter problems while using the equipment, search for the table below for the problem or 
error messages and try the solutions.If the problem persists, turn off the detector and consult your sales 
representative or a distributor. Please refer to the details of the following symptoms or error messages. 
 
 
 
Troubleshooting must be performed by service engineer who is authorized by Vieworks. If 
an unqualified person performs troubleshooting on the system resulting in damaging the 
detector, software or hardware, then the Vieworks or its representative is not responsible 
for the detector repair regardless of remain warranty. For more detailed information, refer 
to <Warranty>. 
8.1.1  Failed to Turn the Detector On 
 
Symptom 
 
Failed to turn on the power of the detector. 
 
Expected Causes 
 
Not installing battery pack well 
 
Dead battery pack 
 
Battery pack or Detector is broken down 
 
Solutions 
1  Install battery pack 
2  Charge battery pack 
3  Check the result after getting rid of the battery pack and connecting the tether cable. 
4  Replace other battery packs and check the result of it. 
5  Replace other Detectors and check the result of it. 
6  Replace corresponding devices. 
8.1.2  The Power Switch of SCU or Status LED is not worked 
 
Symptom 
 
The power switch of Basic SCU or SCU mini is not working. 
 
The status LED of SCU is not responding. 
 
Expected Causes 
 
Power cable is broken down. 
 
Errors in the fuse 
 
Internal circuit is broken down. 
 
Solutions 
1  Check the connection between AC power cable and SCU Basic.