VIVIX-S 1012N User Manual 
4.2 
Product Installation 
4.2.1  Connecting Devices 
 
 
Installation of this equipment should be made by licensed and authorized personnel by 
Vieworks. 
 
This equipment must only be connected to the power with protective earth. 
 
SCU Basic (FXRS-02A) 
 
 
 
1  If you use the DR Trigger mode, connect the one end of generator interface cable to the EXT_INF port 
of SCU, and the other to the X-ray generator. 
 
 
 
If you use the AED mode, a generator interface cable is not needed as the detector 
operates by detecting X-ray automatically. 
 
2  Connect one end of the LAN cable to one of the LAN ports of SCU, and the other to the LAN Card 
Connector of workstation assigned for data transfer. 
3  Connect the one end of the Tether Interface cable to the Port 4 or Port 5 of SCU and the other to the 
detector. 
4  To supply power, connect the AC power cable to the AC power input port of SCU.