When working inside the unit, avoid accessing the battery, other exposed
components, and any item that is protected by the unit's interior plastic
cover. Avoid dropping or spilling anything into the case.
x Using a torx screwdriver with a T25 bit, unscrew and remove the four
components, and any item that is protected by the unit's interior plastic
cover. Avoid dropping or spilling anything into the case.
x Using a torx screwdriver with a T25 bit, unscrew and remove the four
screws that secure the case lid to the case base. Set aside the
screws for later reinstallation.
x The unit is hinged on the left; open the case lid from right to left.
x Bring in the customer-supplied Ethernet cable, power lines, or both
x Bring in the customer-supplied Ethernet cable, power lines, or both
through the customer-installed conduit-hub. If both an Ethernet cable
and fixed DC power are to be used, it is recommended that the cable
occupy one conduit and the power lines occupy the other, following
and fixed DC power are to be used, it is recommended that the cable
occupy one conduit and the power lines occupy the other, following
AHJ requirements.
x Use the diagram below to locate the optional connection ports to be
used: the terminal block, Ethernet cable port, and guides.
Protective cover
RJ45 port
Guides (4x)
Terminal
block
7
Connect power
Next,
choose
only one instruction set below─a or b, but not both─to
connect only one external power source to the RGX.
WARNING: Only one of the following power input ports is permitted to be
connected at a given time: 12V Charger Port, IS Power Port, or 9–30
VDC Terminal Block.
WARNING: Only one of the following power input ports is permitted to be
connected at a given time: 12V Charger Port, IS Power Port, or 9–30
VDC Terminal Block.
a Power supply accessory (ISERTPS or ERTPS)
WARNING: Only one of the following power input ports is permitted to be
connected at a given time: 12V Charger Port, IS Power Port, or 9–30
VDC Terminal Block.
If the RGX will draw operational power from a compatible power supply
connected at a given time: 12V Charger Port, IS Power Port, or 9–30
VDC Terminal Block.
If the RGX will draw operational power from a compatible power supply
accessory (ISERTPS or ERTPS), connect the power supply's IS cable to
the RGX IS Cable Adapter.
Ensure the ISERTPS or ERTPS is installed, connected, and used in
the RGX IS Cable Adapter.
Ensure the ISERTPS or ERTPS is installed, connected, and used in
accordance with the instruction provided in its product manual.
Skip to step 8
.
GND
b Fixed DC power
Document Number: 0000037533
DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS(S):N/A
DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS(S):N/A
Modified: 2018-05-25 13:18:14 EDT
State:In Work
INDUSTRIAL SCIENTIFIC CORPORATION
Title: PRODUCT MANUAL RGX GATEWAY-EN-1
Eng:
Eng:
Drawing No: 17158071-1
Drawn By: Gaygan, Nancy
Created: 2018-05-16 10:24:51 EDT
Rev: 1
Printed By: Kuzmich, Bob
Int: 1.5
Date:May 29, 2018