Industrial Scientific Corporation RGX ユーザーズマニュアル

ページ / 62
 
50 
Conduit-hub Plug Kit installation 
Use only an unused Conduit-
hub Plug Kit from Industrial 
Scientific. 
Do not reinstall any used 
conduit-hub plug components; 
once removed, a plug's seal is 
compromised and other plug 
components may become 
altered or damaged. 
 
 
Inside case 
Hex nut with nylon insert 
Flat washer 
Sealing gasket 
Washer  
Fixed gasket 
 
Outside case 
Washer 
T20 Torx screw 
 
 
 
 
 
Remove the protective paper 
to expose the adhesive side 
of the gasket. 
Inside the case, center and 
place the gasket over the 
conduit-hub plug opening on 
the base.   
From outside of the case, 
place the torx screw into the 
concave side of the washer, 
and place over the base 
opening. 
Inside the case, place the 
next washer—convex side 
down—over the screw and 
fixed gasket. 
 
Follow with the sealing 
gasket, flat washer, and hex 
nut. 
 
 
— — — 
Hold the hex nut in place with a 3/8 ″ open-end wrench and 
turn the T20 torx screw clockwise to tighten.  
Screw torque: 1.27 Nm (180 oz in) 
 
 
 
 
 
 
 
Document Number: 0000037533
DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS(S):N/A
Modified: 2018-05-25 13:18:14 EDT
State:In Work
INDUSTRIAL SCIENTIFIC CORPORATION
Title: PRODUCT MANUAL RGX GATEWAY-EN-1
Eng:
Drawing No: 17158071-1
Drawn By: Gaygan, Nancy
Created: 2018-05-16 10:24:51 EDT
Rev: 1
Printed By: Kuzmich, Bob
Int: 1.5
Date:May 29, 2018