Ortivus AB 410000409 ユーザーズマニュアル

ページ / 47
70
10 Appendix
Database full. No new 
data will be saved for 
this patient.
The limit of 4 hours 
storage of patient 
data has been 
exceeded.
Monitoring of real time data is still possible, however 
no new data will be stored and thus cannot be sent to 
an HWS. A new care session could be started in order 
to store and transmit new data.
MIDA relearn failed.
Poor electrode 
contact.
Check ECG signal quality in Electrode test view. 
Check electrode contact, check if old electrodes, 
check that ECG cable is not damaged. Then try MIDA 
Relearn again.
No communication 
with hospital unit due 
to program error. 
Please contact 
support for more 
information.
Program error.
Contact authorised personnel.
Monitor mode only 
due to program error. 
Please contact 
support for more 
information.
Program error.
Contact authorised personnel.
Facsimile failed.
Fax request did not 
reach hospital unit.
Check communication status and try again.