Appendix 10
10.4.8 Communication
10.4.9 Fax
10.4.10 SOS
Problem
Cause
Action
No contact with a
hospital unit.
Communication is
turned off.
Check if “Communication off” is stated in the
information bar. Turn on communication in settings,
see “Communication settings” on page 59.
Communication has
not been established.
Check if receiver name is in
italics
or in normal font in
information bar, see “Information bar” on page 12. If in
italics, then communication has not been established.
Wait a while and see if changed. Check coverage of
communication network.
Poor coverage of
communication
network.
If there is an antenna/coverage icon in information bar,
check communication network coverage. If bad either
move or change to other communication network if
possible (docked/undocked and reserve channel) see
“Communication” on page 57.
Antenna on the 300m
damaged or not
properly attached.
Check if antenna is damaged or incorrectly attached.
Data not sent to
HWS.
Poor coverage of
communication
network.
If there is an antenna/coverage icon in information bar,
check communication network coverage. If bad either
move or change to other communication network if
possible (docked/undocked and reserve channel), see
“Communication” on page 57.
NOTE! Do not forget to do Retransmit after patient care session in order to
update information at hospital unit for archiving.
Problem
Cause
Action
Facsimile failed.
Different causes
stated in message.
Read message in message log to determine action.
System alarm “Data
not sent”
Indicates that data
has not been sent
due to bad
communication.
If possible, move for better network coverage, or
change to other networks (docked/undocked/reserve)
see “Connecting to a receiver” on page 58. Send the
fax again.
Problem
Cause
Action
No SOS alarm is
received.
Communication is
turned off.
Check if “Communication off” is stated in the
information bar. Turn on communication in settings,
see “Communication settings” on page 59.