Physician Lead Manual
1. Fully insert the lead(s) into the IPG port(s). When the lead is 
properly inserted, the lead will stop and the retention ring will be 
located under the setscrew. 
2. Pass the torque wrench through the slit in the septum located on 
the top of the IPG header and tighten both set screws, one at a 
time, until the torque wrench “clicks,” indicating lock.