Boston Scientific Neuromodulation Corporation PSC5210W ユーザーズマニュアル

ページ / 68
30
Physician Implant Manual
C
AUTION
: Ensure that the lead is fully inserted before tightening the 
setscrew to prevent lead damage.
Note: • If the connector plug is used in port 2-R, it is still necessary 
to tighten the setscrew as described. 
• The wrench is torque-limited and cannot be overtightened.
• Ensure that the lead is fully inserted before tightening the 
setscrew to prevent lead damage.
• If there appears to be an obstruction when inserting the 
lead, lead extension, or connector plug, use the torque 
wrench to loosen (counterclockwise) the setscrew and/or 
gently rotate the lead to help advance the proximal end.
3. Place the IPG in the subcutaneous pocket with “This Side Up” 
facing towards the skin.
4. Coil excess lead or extension under the IPG.
Note: To confirm good connections, check impedances before 
tightening the setscrew.
5. Secure the IPG in the pocket by suturing through the holes in the 
connector.
6. Close and dress the wound(s).