Physician Implant Manual
Note: To save the modifications as a new program, simply select an 
empty [] program slot and press Tinstead of overwriting the 
existing program
.
Options 
With the Precision SCS system, the Pulse Width and Rate parameters 
are generally considered programming “options” in that relatively few 
patients will actually use these settings. By default, Pulse Width and 
Rate are off-limits to the patient Remote Control unless you allow 
access using BionicNavigator software.
Another option, “Restore,” is readily available to all patients. The 
Restore feature allows patients to return a program to the original set-
tings you programmed for them at the initial fitting or at a follow-up. 
To Access the Pulse Width and Rate Options:
1. Press and hold the 
C
button for three seconds (a long button 
press) or until the Pulse Width Area 1 screen appears.
2. As necessary, press the 
C
button normally to cycle to the area 
you wish to adjust. 
3. When the desired area screen is displayed, press S or T to 
increase or decrease the Pulse Width.