Boston Scientific Neuromodulation Corporation PSC5210W ユーザーズマニュアル

ページ / 68
54
Physician Implant Manual
Specifications and Technical Data
Materials
Parameter
Range
Default
Areas (Channels) 
4
Amplitude
0 – 20 mA
a
a. The Precision System includes programmable coverage 
areas with each individual electrode contact limited to 12.7 
mA. A programming interlock is enforced to limit the cover-
age area output current to 20 mA or less. For example, a 
maximum current output of 12.7 mA on a first electrode 
would limit the total summed current output on remaining 
electrodes to 7.3 mA within one coverage area.
0 mA
Rate
2 – 1200 pps
b
b. Only one Area is available if the rate is 130 pps.
40 pps
Width
0 – 1000 µsec
210 µsec
Cycle
0 – 90 min, OFF
OFF
Ramp ON
1 – 10 secs
3 secs
Contacts
1 – 16; +, -, OFF
1 – 16: OFF
Case
Titanium
Header
Epoxy
Strain Relief
Silicone
Size/Volume
55mm x 45mm x 11mm / 22 cm
3
 (including header)