CAS Corporation CASTON3 ユーザーズマニュアル
11
α
Manual Hold Function (Execute when you press HOLD key)
€
Press the HOLD key while weighing an unstable item.
"HoLd" will display and then, the average value of the weight fluctuation will display on
the Weight Display.
€
To escape from this mode, empty the hook or press HOLD key.
Then the scale changes from Hold mode to Normal mode.
(1) How to enter
€
Press the ON/OFF key while pressing the TARE key, and SET mode starts.
(2) Key Usage
€
ZERO : Returns the display to 0.
€
TARE : Increase the set value to 1.
€
SET : Store and move into next menu.
(3) Set Menu(F1 - F5)
€
F1 : Digital Filter Setting
€
F2 : Automatic zero tracking compensation
€
F3 : Weight backup (Power on actual weight)
NOTE. If you press SET key after F3 setting, it will display "bye" and enter to Normal mode.
€
F4 : Zero key operation range set
€
F5 : Option select
10. SET MODE
Set Value
Description
F1- 1
In high speed (Less vibration)
F1- 4
In normal speed
F1- 9
Very slowly (Much vibration)
Set Value
Description
F2- 0
No compensation
F2- 4
Compensation for gradual change below 2 divisions.
F2- 9
Compensation for gradual change below 4.5 divisions.
Set Value
Description
F3- 0
Weight backup is OFF
F3- 1
Weight backup is ON
Set Value
Description
F4- 0
±
2% : Zero key operation within
±
2% of maximum weight.
F4- 1
±
10% : Zero key operation within
±
10% of maximum weight.