PART II. Service Manual 
 
1417WGC/WCC 
94 
8. 
“Acquiring bright frame” is closed after radiation is detected, and the program will show if 
the detected radiation is within acceptable range. 
 
When PASS is displayed 
When FAIL is displayed 
Shoot X-ray with same technique when 
“Acquiring bright frame” is popped up. 
Adjust technique to get acceptable value and 
shoot again. 
9. 
For each Calibration point, four images must successfully be acquired. After successfully 
doing so for every point, the Calibration process is complete. 
 
10. 
Click "OK" to move to the next step.