Rayence Co. Ltd. RY1417WHD ユーザーズマニュアル
PART II. Service Manual
1417WGC/WCC
94
8.
“Acquiring bright frame” is closed after radiation is detected, and the program will show if
the detected radiation is within acceptable range.
the detected radiation is within acceptable range.
When PASS is displayed
When FAIL is displayed
Shoot X-ray with same technique when
“Acquiring bright frame” is popped up.
“Acquiring bright frame” is popped up.
Adjust technique to get acceptable value and
shoot again.
shoot again.
9.
For each Calibration point, four images must successfully be acquired. After successfully
doing so for every point, the Calibration process is complete.
doing so for every point, the Calibration process is complete.
10.
Click "OK" to move to the next step.