Sensors & Software Inc. PE-PRO-TLF ユーザーズマニュアル

ページ / 34
pulseEKKO PRO TLF Transmitter
4-Care and Maintenance
13
All battery compartments should be checked periodically for proper operation.  Take a known
good battery and install into the four battery compartments.  Ensure the power light activates with
no intermittent operation.  If the power light will not activate look in the battery compartment and
check the battery contact for signs of excessive stress.  Also check the condition of the 2A fuse
wiring.  If the problem persists, remove the six screws holding the battery plate in place and lift
upwards.  On the bottom of the battery plate you will see a small circuit board that extends the
battery input to the power switch.  Check the connections to and from this board.  Please contact
Sensors & Software Inc.  should all the above fail to resolve the problem.  
4.6
Fiber Optics Cables
These cables can stand a great deal of abrasive wear but the user should avoid sharp bends in
the cable, pulling the cable at right angles close to the fiber optics connectors, or allowing the
unprotected ends to get dirty or scratched.  Erratic timezero jumps usually indicate a cracked
cable.  If problems are encountered with the fiber optics cables, repair is very straightforward and
is outlined in detail below.  The best policy is to carry spares.  Short fiber optic cables are located
inside each of the units.  These short cables may need cleaning and polishing from time to time.