Given Imaging Limited CAPDH-2 ユーザーズマニュアル

ページ / 250
PillCam Equipment Maintenance
PillCam Recorder Maintenance
203
PillCam Recorder DR2
Charging
Make sure the PillCam recorder is at least 80% charged (eight or more battery bars displayed on the 
battery icon) for SB and COLON procedures, and at least 50% charged (five or more bars displayed on 
the battery icon) for ESO procedures.
The cradle has the following LEDs:
To charge the PillCam recorder:
1.
Plug the power cable into the cradle and plug the power cable into the wall outlet.
All three LEDs turn on for a self-test that takes 5 seconds. After 5 seconds, all LEDs turn off and 
the cradle is ready for use.
If during the self-test the red LED blinks, the cradle is faulty. Contact Given Imaging Customer 
Support.
2.
Insert the PillCam recorder or its Li-Ion battery with its adaptor into the cradle.
All three LEDs of the cradle blink for 4 seconds. When the orange LED is on, charging has started.
As soon as the PillCam recorder or its battery pack are fully charged, the green LED turns on, and 
the orange LED turns off. 
3.
Leave the PillCam recorder in its cradle until the next examination. 
LED
Status
Battery pack is...
Green 
On
ready for use
Orange
On
charging
Blinking
discharging
Red
On
faulty (malfunctioning)
Adaptor with 
battery
Green, orange, 
and red LEDs