Preparing for PillCam Capsule Endoscopy
Fitting Equipment on the Patient
53
5.
With the sensor array connector at the patient's side, lay the prepared sensor array on the patient's 
abdomen and match the letters and colors on each sensor wire to the letters and colors on the 
Sensor Location Guide. 
6.
To attach each sensor, remove the protective backing from its adhesive sleeve.
If you are using a sensor array with a downlink loop, adjust the downlink loop to remove excess 
cable so that it fits closely to the patient's body. To prevent damage to the wires do not forcibly 
bend the downlink loop in any way. 
7.
When the patient gets dressed, make sure that the sensor array connector remains outside of the 
patient's clothing and make sure that the patient is not uncomfortable with the equipment.
֠
֠
֠
֠
Note
Since the sensors are placed on the body according to anatomical reference 
points, the distances between sensors may vary from patient to patient.