Given Imaging Limited GIVENRECORDER ユーザーズマニュアル
PillCam Platform
4
Chapter 1
The battery of the DataRecorder 2 is charged in the cradle either with its adaptor or while inside
the DataRecorder 2.
the DataRecorder 2.
The DataRecorder 2 is ready for operation when its battery is charged and the SensorArray is
connected. When the DataRecorder 2 is on, it starts recording as soon as a signal is received from
any PillCam video capsule. When the capsule LED blinks, the DataRecorder 2 is receiving data.
When the signal from the PillCam video capsule is too weak, the LED does not blink.
connected. When the DataRecorder 2 is on, it starts recording as soon as a signal is received from
any PillCam video capsule. When the capsule LED blinks, the DataRecorder 2 is receiving data.
When the signal from the PillCam video capsule is too weak, the LED does not blink.
DataRecorder 2 Automatic Shutdown
After the DataRecorder 2 has been initialized with patient data, it goes into a standby mode when
removed from its cradle and starts recording as soon as a signal is received from any PillCam
video capsule. If no signal is received, the DataRecorder automatically shuts down after 90
minutes. This feature ensures that the DataRecorder 2 preserves sufficient battery power to record
a complete study.
removed from its cradle and starts recording as soon as a signal is received from any PillCam
video capsule. If no signal is received, the DataRecorder automatically shuts down after 90
minutes. This feature ensures that the DataRecorder 2 preserves sufficient battery power to record
a complete study.
Capsule LED
DataRecorder LED