Given Imaging Limited SMARTPILL ユーザーズマニュアル

ページ / 120
Appendix: Electrical Safety 
x history of gastric bezoar 
x swallowing disorders 
x suspected or known strictures, fistulas, or physiological/mechanical GI 
obstruction 
x GI surgery within the past 3 months 
x severe dysphagia to food or pills 
x Crohn’s disease or diverticulitis 
x implanted or portable electro-mechanical medical device such as a cardiac 
pacemaker, defibrillator or infusion pump 
x younger than 18 years old. 
Data transmission from the capsule to the data receiver is influenced by patient BMI. 
Significant data dropout can occur in severely obese patients (>40 BMI).   
Restricted Use 
Caution 
The SmartPill GI Monitoring System equipment is not suitable for use in the presence of flammable anesthetic 
mixture with air or with oxygen or nitrous oxide. 
Not for use with oxygen or oxygen-enriched atmospheres. 
Storage 
Store SmartPill GI Monitoring System components and capsules at ambient room 
temperature (-15–40ºC) and humidity (rH 30–90%). 
Caution 
Do not expose the capsules to UV light. UV light can permanently damage the pH sensor. 
Caution 
Do not store capsules within 30cm (1 foot) of the activation fixture. The fixture’s magnetic field could 
inadvertently activate the capsules. 
Power Requirements 
System Computer 
110/220 VAC, 50–60 Hz 
May require the use of a 
power plug adapter 
Docking Station 
110/220 VAC, 50–60 Hz 
May require the use of a 
power plug adapter 
Capsule  
3.1 VDC self-contained 
batteries 
none 
Data Receiver 
8.4 VDC, self-contained, 
none