    Fusion Wireless Recorder 
 
Setup Steps 
This manual is designed to allow a technician to follow the instructions 
page by page to setup the Fusion Recorder.  Here is the general layout: 
1.  Connect leads and electrodes to patient. 
2.  Prepare Recorder for recording. 
A.  Choose/Setup program options you want to use.  
B.  Erase all previous data. 
3.  Connect Patient Cable to Recorder. 
 
Electrode Application and Placement 
For each electrode lead wire: 
1.  Snap the electrode onto the lead wire. 
2.  Remove the protective backing from the adhesive side of the 
electrode. 
3.  Apply the electrode to the patient’s skin per Electrode Placement 
diagram in this manual or as instructed by the physician.  
Notes: 
A.  It is recommended that trained medical personnel instruct the 
patient in the proper application of electrodes. 
B.  Use good quality long term electrodes. Braemar recommends 
the use of low impedance Holter electrodes. Instruct patient to 
apply fresh electrodes regularly. (Usually on a daily basis.) 
C.  Proper preparation of the patient's skin is absolutely essential 
for obtaining a quality ECG recording.  The skin surface where 
the electrodes will be placed should be cleaned with alcohol, 
allowed to dry, and abraeded. 
D.  Any loose electrode needs to be replaced. 
 
 
 
 
      
    Fusion Wireless Recorder 
 
1/2/3 Channel Electrode Placement 
 
This is a typical electrode placement. Refer to Analysis System software 
and the physician for recommended positioning.   
 
 
 
 
 
 1, 2, and 3 Channel Electrode Placement 
 
 
Channel 1  = Red (1+), White (1-),  
Green (ground (RL)) 
Channel 2   = Red (1+), White (1-/2-), Black (2+), 
Green (ground (RL)) 
Channel 3  = Brown (1+), Red (1-/2-), Black (2+,3+), 
White (3-), Green (ground (RL))