Biolase Inc. EPIC ユーザーズマニュアル
26 |
P a g e
5 4 0 0 3 2 1 R e v . X : D R A F T
PROCEDURES Button
The
EPIC 10
has 20 pre-set procedures, 14 with pre-programmed procedural presets and 6
empty slots for custom pre-sets. All of them can be customized to your preference.
In order to customize parameters for the particular clinical procedure:
In order to customize parameters for the particular clinical procedure:
Adjust parameters on the main Pre-Programmed Procedure menu
Select PROCEDURES Mode
Press and hold the selected Procedure for 2 seconds. Parameters will be changed
Select PROCEDURES Mode
Press and hold the selected Procedure for 2 seconds. Parameters will be changed
and saved for that Procedure (you will hear a beeping sound when the adjusted
settings are saved).
settings are saved).
Figure 25
SETTINGS
Screen
By pressing the Settings icon on the main screen, you can access several system settings:
Language
Selection
Aiming Beam (5 levels of brightness adjustment)
Beep Sound (3 levels of sound adjustment)
Service Mode (accessible only by authorized Biolase Service Representatives)
Restore Factory Default Settings
Beep Sound (3 levels of sound adjustment)
Service Mode (accessible only by authorized Biolase Service Representatives)
Restore Factory Default Settings
Turn the Unit OFF
Place handpiece back on handpiece holder.
CAUTION:
Verify that fiber optic tubing assembly is not twisted once the handpiece
is returned to the holder. The fiber may break if it is twisted.
is returned to the holder. The fiber may break if it is twisted.
Press the CONTROL button to turn display OFF.
Switch the Power Switch at the rear of the base of the console to OFF position, if the
laser system will not be used for a long period of time.
Wind the fiber cable onto the fiber spool around the base of the console.