Biolase Inc. EPIC ユーザーズマニュアル

ページ / 45
26 | 
P a g e
                                                                             
5 4 0 0 3 2 1 R e v .   X :   D R A F T
 
 
PROCEDURES Button 
The 
EPIC 10
 has 20 pre-set procedures, 14 with pre-programmed procedural presets and 6 
empty slots for custom pre-sets. All of them can be customized to your preference.  
In order to customize parameters for the particular clinical procedure:  
  Adjust parameters on the main Pre-Programmed Procedure menu 
  Select PROCEDURES Mode  
  Press and hold the selected Procedure for 2 seconds. Parameters will be changed 
and saved for that Procedure (you will hear a beeping sound when the adjusted 
settings are saved). 
 
 
                                                                     Figure 25
 
 
 
 
 
 
 
     
 
SETTINGS
 
Screen 
By pressing the Settings icon on the main screen, you can access several system settings: 
 Language 
Selection 
  Aiming Beam (5 levels of brightness adjustment) 
  Beep Sound (3 levels of sound adjustment) 
  Service Mode (accessible only by authorized Biolase Service Representatives) 
  Restore Factory Default Settings 
 
Turn the Unit OFF 
  Place handpiece back on handpiece holder. 
 
CAUTION:
  
Verify that fiber optic tubing assembly is not twisted once the handpiece 
is returned to the holder. The fiber may break if it is twisted.
 
 
  Press the CONTROL button to turn display OFF. 
 
  Switch the Power Switch at the rear of the base of the console to OFF position, if the 
laser system will not be used for a long period of time. 
 
  Wind the fiber cable onto the fiber spool around the base of the console.