Biolase Inc. EPIC ユーザーズマニュアル

ページ / 45
32 | 
P a g e
                                                                             
5 4 0 0 3 2 1 R e v .   X :   D R A F T
 
 
PRE-PROGRAMMED SETTINGS 
 
PROCDURE 
NAME 
MODE 
PEAK 
POWER 
AVERAGE  
POWER 
PULSE  
INTERVAL 
PULSE 
LENGTH 
DUTY  
CYCLE 
TIP            
TYPE 
INITIATION 
1 GINGIVECTOMY  CP0 
5.0W 
1.0W 
0.04 ms 
0.01 ms 
20% 
E4 
YES
 
2 TROUGHING  CP2 
2.0 W 
1.0 W 
1.0 ms 
1.0 ms 
50% 
E4 
YES
 
3 CURETTAGE  CP1 
2.4 W 
0.8 W 
0.2
 ms 
0.
1
 ms 
30% 
E4 
YES
 
4 EXCISION  CP1 
2.7 W 
0.9 W 
0.2
 ms 
0.
1
 ms 
30% 
E4 
YES
 
5 FRENECTOMY  CP2 
2.0 W 
1.0 W 
1.0 ms 
1.0 ms 
50% 
E4 
YES
 
IMPLANT  
RECOVERY 
CP2 
2.5 W 
1.25 W 
1.0 ms 
1.0 ms 
50% 
E4 
YES
 
7 PERIO 
POCKETS  CP2 
1.6 W 
0.8 W 
1.0 ms 
1.0 ms 
50% 
E3 
NO
 
PULPOTOMY(
*
CW 
0.10W 0.10W 
N/A
 
N/A
 
N/A
 
E4
 
YES
 
CROWN 
LENGTHENING 
CP1 
2.7 W 
0.9 W 
0.2
 ms 
0.
1
 ms 
30% 
E4 
YES
 
10 
INFECTED 
POCKETS 
CP2 
1.6 W 
0.8 W 
1.0 ms 
1.0 ms 
50% 
E4 
YES
 
11 
ENDO(
*
CW 
0.10 W 
0.10 W 
N/A
 
N/A
 
N/A
 
E2
 
NO
 
12 HEMOSTASIS  CW 
0.50W 0.50W 
N/A
 
N/A
 
N/A
 
E4
 
YES
 
13 APHTHOUS 
ULCERS  CW 
0.70W 0.70W 
N/A
 
N/A
 
N/A
 
E4
 
NO
 
14 
EXPOSURE OF 
UNERUPTED TEETH 
CP2 
1.80W 0.90W 
1.0ms 
1.0ms 
50% 
E4 
YES
 
15 
CUSTOM 1 - 6 
 
 
 
 
 
 
 
 
 
 
     (
*
) Minimum defaults provided for user setting of Endodontic Procedures such as Pulpotomy and 
Pulpotomy as an adjunct to root canal therapy.